Published September 11, 2007
| Version v1
Journal article
The statistics of multi-photoelectron pulse-height distributions
Description
Multi-photoelectron pulse-height distributions feature in the scientific literature on light detectors, and yet there appears to have been no attempt to describe the statistics of such spectra. Formulas for pdfs are derived, assuming Poisson statistics, but more importantly for single-electron response curves of any description-analytical or experimental. The pdf generated by a pulsed light source, detected by a PMT of high d1 gain, agrees with statistical predictions
Availability note (English)
Available from http://dx.doi.org/10.1016/j.nima.2007.06.021Additional details
Identifiers
- DOI
- 10.1016/j.nima.2007.06.021;
- PII
- S0168-9002(07)01218-1;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 579
- Journal Issue
- 3
- Journal Page Range
- p. 967-972
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 39061013
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- DIAGRAMS; DISTRIBUTION; ELECTRONS; FORECASTING; GAIN; HEIGHT; LIGHT SOURCES; PHOTOMULTIPLIERS; PULSES; RESOLUTION; SPECTRA; STATISTICS; VISIBLE RADIATION
- Descriptors DEC
- AMPLIFICATION; DIMENSIONS; ELECTROMAGNETIC RADIATION; ELEMENTARY PARTICLES; FERMIONS; INFORMATION; LEPTONS; MATHEMATICS; PHOTOTUBES; RADIATION SOURCES; RADIATIONS
Optional Information