Published 2011 | Version v1
Miscellaneous

Preliminary validation of a Monte Carlo model for IMRT fields

  • 1. Australian Radiation Protection and Nuclear Safety Agency (ARPANSA), Melbourne, Victoria, Australia (Australia)
  • 2. Department of Medical Physics, Royal Adelaide Hospital, Adelaide, South Australia (Australia)
  • 3. School of Chemistry and Physics, University of Adelaide, Adelaide, South Australia, Australia (Australia)

Description

Full text: A Monte Carlo model of an Elekta linac, validated for medium to large (10-30 cm) symmetric fields, has been investigated for small, irregular and asymmetric fields suitable for IMRT treatments. The model has been validated with field segments using radiochromic film in solid water. The modelled positions of the multileaf collimator (MLC) leaves have been validated using EBT film, In the model, electrons with a narrow energy spectrum are incident on the target and all components of the linac head are included. The MLC is modelled using the EGSnrc MLCE component module. For the validation, a number of single complex IMRT segments with dimensions approximately 1-8 cm were delivered to film in solid water (see Fig, I), The same segments were modelled using EGSnrc by adjusting the MLC leaf positions in the model validated for 10 cm symmetric fields. Dose distributions along the centre of each MLC leaf as determined by both methods were compared. A picket fence test was also performed to confirm the MLC leaf positions. 95% of the points in the modelled dose distribution along the leaf axis agree with the film measurement to within 1%/1 mm for dose difference and distance to agreement. Areas of most deviation occur in the penumbra region. A system has been developed to calculate the MLC leaf positions in the model for any planned field size.

Part of:
Abstract EPSM-ABEC 2011 Conference

Additional details

Publishing Information

Publisher
Australasian College of Physical Scientists and Engineers in Medicine
Imprint Place
Darwin (Australia)
Imprint Title
Abstract EPSM-ABEC 2011 Conference
Imprint Pagination
79 p.
Journal Page Range
p. 602

Conference

Title
The Engineering and Physical Sciences in Medicine and the Australian Biomedical Engineering Conference
Acronym
EPSM-ABEC 2011
Dates
14-18 Aug 2011
Place
Darwin (Australia)

INIS

Country of Publication
Australia
Country of Input or Organization
Australia
INIS RN
44026382
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
CALCULATION METHODS; COLLIMATORS; CT-GUIDED RADIOTHERAPY; MONTE CARLO METHOD; NUCLEAR MEDICINE; RADIATION DOSE DISTRIBUTIONS; RADIOLOGY; RADIOTHERAPY; SPATIAL DOSE DISTRIBUTIONS
Descriptors DEC
CALCULATION METHODS; MEDICINE; NUCLEAR MEDICINE; RADIATION DOSE DISTRIBUTIONS; RADIOLOGY; RADIOTHERAPY; THERAPY

Optional Information