Published March 2001 | Version v1
Report Open

Quality control and patient doses from X-ray examinations in some hospitals in Thailand

  • 1. Ministry of Public Health, Nonthaburi (Thailand)

Description

Quality control measurements on 203 diagnostic X-ray units were carried out in 126 hospitals in the central region of Thailand during 1998-2000. The measurements consisted of tube voltage, half-value layer (HVL), exposure time, radiation output, beam alignment, light beam diaphragm and entrance surface dose (ESD) in four common radiographic procedures, namely adult chest PA, adult mass chest PA in mobile bus unit, abdomen AP and mammography (cephalo caudal view). ESD measurements of 320 examinations were performed using parallel plate ionization chamber and Keithley model 35050A Dosimeter on 192 X-ray units (conventional and mobile) and 11 mammography units. The analysis of test results showed that: 1) 92% had X-ray tube voltage within the tolerance limit of 10% and HVL 3.03 mmAl ± 0.55 SD at 80 kVp measured; 2) 90% had exposure time within tolerance limit of 10%; 3) 86% and 98% had acceptable beam alignment and light beam diaphragm; 4) 95% had radiation output > 25 μGy at 1 m. for true 80 KVp; 5) It was found that the ESD values were adult chest (PA) varied from 0.18 mGy to 1.17 mGy (mean 0.2021 ± 0.2218 SD), adult mass chest (PA) varied from 0.043 mGy to 1.03 mGy (mean 0.2935 ± 0.2195 SD), abdomen (AP) varied from 0.302 mGy to 6.30 mGy (mean 2.177 ± 1.4818 SD), mammography (cephalo caudal view) varied from 3.49 mGy to 12.21 mGy (mean 7.788 ± 2.9896 SD). Further surveys are necessary and are being done to include measurements of image quality and for propagation of quality assurance activities in Thailand so as to reduce patient doses while maintaining the image quality. (author)

Files

32039835.pdf

Files (164.6 kB)

Name Size Download all
md5:e754a177c59ddc875ea06ec7269aa3e9
164.6 kB Preview Download
Part of:
Radiological protection of patients in diagnostic and interventional radiology, nuclear medicine and radiotherapy. Contributed papers

Additional details

Publishing Information

Imprint Title
Radiological protection of patients in diagnostic and interventional radiology, nuclear medicine and radiotherapy. Contributed papers
Imprint Pagination
916 p.
Journal Issue
no. 7/P
Series
C and S papers series
Journal Page Range
p. 99-103
ISSN
1563-0153
Report number
IAEA-CSP--7/P

Conference

Title
International conference on radiological protection of patients in diagnostic and interventional radiology, nuclear medicine and radiotherapy
Dates
26-30 Mar 2001
Place
Malaga (Spain)

INIS

Country of Publication
International Atomic Energy Agency (IAEA)
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
32039835
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; DOSE LIMITS; DOSIMETRY; ELECTRIC POTENTIAL; HOSPITALS; IMAGE PROCESSING; PATIENTS; QUALITY CONTROL; RADIATION DOSES; RADIATION PROTECTION
Descriptors DEC
BUILDINGS; CONTROL; DIAGNOSTIC TECHNIQUES; DOSES; MEDICAL ESTABLISHMENTS; MEDICINE; NUCLEAR MEDICINE; PROCESSING; RADIOLOGY; SAFETY STANDARDS; STANDARDS

Optional Information

Notes
5 refs, 5 figs, 3 tabs
Secondary number(s)
IAEA-CN--85-288