Published February 1980 | Version v1
Journal article

An electro-optical ion detector for spark source mass spectrometry

  • 1. Oak Ridge National Lab., TN (USA). Analytical Chemistry Div.
  • 2. Tennessee Univ., Knoxville (USA). Dept. of Chemistry

Description

An ion detector for spark cource mass spectrometry is described which utilizes a channel electron multiplier array in the chevron configuration with phosphor screen display. An optical flange positions the detector in the focal plane of an AEI-MS702R mass spectrometer. The photographic plate detection mode of the spectrometer is also available with a minimum of change-over time. Preliminary results are shown for several modes of electronic data collection. (orig.)

Additional details

Publishing Information

Journal Title
Int. J. Mass Spectrom. Ion Phys.
Journal Volume
33
Journal Issue
1
Series
Int. J. Mass Spectrom. Ion Phys.
Journal Page Range
45-55