Published May 20, 1999
| Version v1
Report
ENRAF annulus leak detector development test; TOPICAL
Description
ENRAF 854 ATG level detectors are normally used for measuring and tracking waste levels in the Hanford waste tanks. Since instrument technicians and operators are familiar with these gauges, it has been suggested that ENRAF 854's be used to alleviate problems with the existing annulus leak detector devices. The purpose of this test is to determine the following: (1) The liquid level at which the gauge will report a change in level, given the displacer initially resting on the bottom of a container; and (2) Whether the high alarm level can be set at 0.25 inch above the bottom of the container. This test and its results will be applicable to ENRAF 854 ATG level detectors used as annulus leak detectors
Availability note (English)
Available from OSTI as DE00782315; www.osti.gov/servlets/purl/782315-QByKOo/native/; www.osti.gov/servlets/purl/782315-Ff92nE/webviewable/Additional details
Identifiers
Publishing Information
- Imprint Pagination
- 9 p.
- Report number
- HNF--4033-REV0
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 33064120
- Subject category
- S12: MANAGEMENT OF RADIOACTIVE WASTES, AND NON-RADIOACTIVE WASTES FROM NUCLEAR FACILITIES;
- Resource subtype / Literary indicator
- Non-conventional Literature
- Descriptors DEI
- ANNULAR SPACE; LEAK DETECTORS; LEVEL INDICATORS; PERFORMANCE TESTING; RADIOACTIVE WASTE FACILITIES; RADIOACTIVE WASTE STORAGE; STORAGE FACILITIES
- Descriptors DEC
- CONFIGURATION; MANAGEMENT; MEASURING INSTRUMENTS; NUCLEAR FACILITIES; RADIOACTIVE WASTE MANAGEMENT; SPACE; STORAGE; TESTING; WASTE MANAGEMENT; WASTE STORAGE
Optional Information
- Contract/Grant/Project number
- AC06-96RL13200
- Funding organization
- US DOE Office of Environmental Management (United States)