Concept and equipment of high resolution time-of-flight trigger

Description

The time-of-flight trigger system for MASPIC two-arm magnet spectrometer is described. The scintillation counter signals that are timing to charged particles hitting into counter are used for selection as input signals. The concept is based on the use of the coincidence method with programmable coincidence resolving time, delays and turning of channels. The PRAVETS-16 controls the system. The particle time-of flight is fixed by timing setup with a 300 ps accuracy. The coincidence circuit resolving time range can be fixed from 1 to 15 ns with pitch ∼ 65 ps. The delays are installed with the same accuracy. The system provides the effective separation of particles that time-of-flight difference exceed 1 ns. The work time is less than 50 ns. 8 refs.; 2 figs

Availability note (English)

MF available from INIS under the Report Number.

Files

21032440.pdf

Files (230.4 kB)

Name Size Download all
md5:9c198fd680a35473765cce73252b3f00
230.4 kB Preview Download

System files (22.1 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Организация и аппаратурная реализация триггера по времени пролета с высоким разрешением

Publishing Information

Imprint Pagination
6 p.
Report number
JINR--13-88-437