Published 2017 | Version v1
Book

Monte Carlo calculation of the efficiency calibration curve for HPGe detectors

  • 1. Atomic Minerals Directorate for Exploration and Research, Bengaluru (India)
  • 2. Bangalore University, Bengaluru (India)
  • 3. Atomic Minerals Directorate for Exploration and Research, Hyderabad (India)

Description

In this work, efficiency of the HPGe detector present at Physics lab, AMD, Bengaluru was simulated using MCNP4C2. During the work, efficiency curves for 0 to 2.8 MeV energy were generated for different weight and densities of 0.26% U3O8 and simulated spectra were compared with experimentally generated spectra. Quartz powder was used as matrix since it is generally used to dilute the uranium ore because of its negligible radioactivity

Part of:
Proceedings of the national conference on radiation physics

Additional details

Publishing Information

Publisher
Bangalore University
Imprint Place
Bengaluru (India)
Imprint Title
Proceedings of the national conference on radiation physics
Imprint Pagination
316 p.
Journal Page Range
p. 161-165

Conference

Title
national conference on radiation physics
Acronym
NCRP-2017
Dates
23-24 Nov 2017
Place
Bengaluru (India)

Optional Information

Notes
3 refs., 6 figs., 1 tab.