Published 2010 | Version v1
Journal article

A phantom for investigation of tumour signal and noise in PET reconstruction with various smoothing filters: Experiments and comparisons with simulated intensity diffusion

  • 1. The Interventional Centre, Oslo Univ. Hospital and Inst. of Hospital Medicine, Oslo (Norway)
  • 2. Dept. of Radiology and Nuclear Medicine, Oslo Univ. Hospital and Inst. of Hospital Medicine, Univ. of Oslo, Oslo (Norway)

Description

An abdomen phantom, containing an idealised fillable liver volume was developed. A fast settling gel was used to produce a row of simulated tumours as well as normal surrounding tissue, each of these components containing different concentrations of 18F-FDG. The magnitude of the signal above normal tissue intensity as well as the noise was obtained after reconstructions with Ordered Subset expectation-maximization (OSEM) and various Gaussian smoothing filters. Signals derived by 'simulated intensity diffusion' agreed well with experimental results. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncq051

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
139
Journal Issue
1-3
Journal Page Range
p. 191-194
ISSN
0144-8420

Conference

Title
3. Malmoe Conference on Medical Imaging - Optimisation in X-ray and molecular imaging
Dates
25-27 Jun 2009
Place
Malmoe (Sweden)

Optional Information

Notes
3 refs