Published December 1, 2006 | Version v1
Journal article

Anomalous behavior of LiF:Mg,Cu,P (TLD100H) irradiated with electron beams

  • 1. Dipartimento di Fisica e Tecnologie Relative, Viale delle Scienze, Pad. 18, 90128 Palermo (Italy)
  • 2. Unita CNISM, Viale delle Scienze, Pad. 18, 90128 Palermo (Italy)
  • 3. Dipartimento Farmacochimico, Tossicologico e Biologico, Via Archirafi 32, 90123 Palermo (Italy)

Description

Thermoluminescent dosimeters of LiF:Mg,Cu,P (TLD100H) were exposed to 60Co γ-ray, protons, high energy electrons and the glow curves were analyzed. We observed an unusual behavior of the TL response of dosimeters to 7 and 14MeV electron beam. In fact the relative amplitude between two of the five peaks (2oand 3opeaks) assumes values smaller than the unity for doses up to 5Gy and values greater than the unity for doses equal to or greater than 10Gy. A possible explanation of this behavior is given adopting a model of the formation of the peaks 2oand 3oand examining the different ways of energy release in matter for each beam. To perform this analysis we deconvoluted the glow curve adopting general order kinetic (GOK) equation for the description of the peak shape, determining the number of peaks and their trapping parameter (E,s,n0)

Additional details

Identifiers

DOI
10.1016/j.nima.2006.08.082;
PII
S0168-9002(06)01512-9;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
568
Journal Issue
2
Journal Page Range
p. 784-791
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2006 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.