Description

The energy resolution of a superconducting tunnel junction particle detector was significantly improved by the addition of a metal mesh infrared-blocking filter. The energy resolution was 138-215 eV full width at half maximum for 1.5-3.0 keV Ar ions. The infrared-blocking filter reduces the loss of energy resolution due to infrared radiation at room temperature

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2008.07.078