Published December 2013 | Version v1
Journal article

Technetium-99m eluate quality affects in vitro red blood cell labeling of healthy human volunteers blood: An university teaching hospital experience

  • 1. Department of Nuclear Medicine of SOAHS, Manipal University, Manipal, Karnataka (India)

Description

Purpose of the present study is to report and disseminate an initial experience gained with the 99mTechnitium sodium pertechnetate eluate; obtained after second elution of the 99Moly-99mTc column generator within 3-4 h on the same day; influences the labeling efficiency of 99mTc labeled RBC while performing by in vitro method in the hospital pharmacy laboratory. Eluate from 99Moly- 99mTc column generator was used consisting radioactive concentration of 50 μCi in 1 ml of normal young healthy human volunteers (n = 27, male =14 and female =13) has not consumed any interfering synthetic chemicals/drugs blood collected in the heparinised syringe after taking informed consent duly approved by Institutional Ethics committee. SnCl2.2H2O used as reducing agent for 99mTc of 2 μg/ml of concentration with mixing, incubating 20-30 min, separating packed red cells, washing with 0.9% saline and then counted in the calibrated Gamma ray Spectrophotometer. Percentage labeling efficiency of 99mTc-RBC was calculated. Statistical variation Mean ± SD was calculated. Labeling efficiency by in vitro method of RBC collected from human volunteers blood with the 99mTc was showing poor percentage of 56.5 ± 3.8 as compared to the expected standard percentage binding of ≥90%. Result analysis reflecting that 99mTc04 − eluate used for RBC labeling may be containing more ground Technetium (99Tc (ground)) that means ratio of 99mTc to 99Tc (g) may be very low, which has interfered with the 99mTc (Gamma emitter 140 keV). Limitation of the study is that 99Tc (g) in the eluate from the column generator has to be quantified with the suitable and feasible method at different intervals to provide more evidence. Study concludes that 99mTco4 − eluted for the red blood cell labeling should be fresh and should not be used if the generator has been eluted earlier within 3-4 h. (author)

Additional details

Publishing Information

Journal Title
Indian Journal of Nuclear Medicine
Journal Volume
28
Journal Issue
5,suppl
Journal Page Range
p. 50
ISSN
0972-3919