Published October 10, 1989 | Version v1
Journal article

An updated system of electronic modules for X-ray spectrometers with cooled semiconductor detectors

  • 1. Ural'skij Politekhnicheskij Inst., Sverdlovsk (USSR)

Description

An updated system of program-controlled analog and analog-to-digital modules for X-ray spectrometers with semiconductor detectors is reported. Ways to increase the count rate capacity of pulse-drain feedback charge-sensitive preamplifiers to 106 cps for EX=5.9 keV are considered in detail. A new analog processor model AP-007 incorporating Kandiah's cusp-shaper is described. Some problems of further improvements in the modular system are discussed. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research, Section A
Journal Volume
282
Journal Issue
2/3
Series
Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
734-738
ISSN
0168-9002
CODEN
NIMAE

Conference

Title
8. international conference on synchrotron radiation utilization (SR-8).
Dates
18-22 Aug 1988.
Place