Description

A single collimator version of a proposed PG-SPECT system was manufactured and experimentally tested. Combining this experimental data with Monte Carlo simulation, the viability of Ge and CdTe semiconductors detectors was calculated. It was determined that the best detector of the ones compared would be a CdTe detector of 2–3 mm, aided by the benefit of adding a Compton-suppression anti-coincidence timing detector. - Highlights: • A single collimator version of a PG-SPECT system was experimentally tested. • The rate of 478 keV and 2223 keV gamma-rays passing through the aperture was measured. • Monte Carlo simulation was used to evaluate various semiconductor detectors. • Of the detectors simulated, a CdTe detector of thickness 2–3 mm is the best

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2013.11.135

Additional details

Identifiers

DOI
10.1016/j.apradiso.2013.11.135;
PII
S0969-8043(13)00585-X;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
88
Journal Issue
Complete
Journal Page Range
p. 167-170
ISSN
0969-8043
CODEN
ARISEF

Conference

Title
15. international congress on neutron capture therapy impact of a new radiotherapy against cancer
Dates
23 Aug 2005
Place
Tsukuba (Japan)

Optional Information

Copyright
Copyright (c) 2013 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.