Published December 2013 | Version v1
Journal article

Development of a pinhole gamma camera for environmental monitoring

  • 1. Department of Radiological Science, Dongseo University, Busan 617-716 (Korea, Republic of)
  • 2. Institute of Health Science, Yonsei University, Wonju 220-710 (Korea, Republic of)
  • 3. Department of Radiological Science, College of Health Science, Yonsei University, 234 Maeji, Heungup, Wonju, Kangwon-Do 220-710 (Korea, Republic of)
  • 4. Korea Institute of Nuclear Nonproliferation and Control, Daejeon 305-348 (Korea, Republic of)

Description

The collimator design for a nuclear monitoring system should be considered differently from the collimator design for medical environments because it has to be used in high-energy radiation environments. The purpose of this study was to determine the optimum pinhole design and to evaluate its performance for acquiring good-quality image in a high-energy radiation field. Simulations using the Geant4 Application for Tomographic Emission (GATE) were performed to model the pinhole gamma camera system. The gamma camera consists of a pyramid-shaped lead collimator with a tungsten pinhole insert, and a CsI(Tl) scintillation crystal with thickness of 6.0 mm and area of 50.0 mm × 50.0 mm. The acceptance angle of the pinhole collimator and the distance from pinhole to scintillator crystal were set to 45° and 60 mm, respectively. The intrinsic spatial resolution and sensitivity were simulated by changing the pinhole diameter and channel height. The point source was located 60 mm above the center of the pinhole, and the transmitted image was estimated for pinhole diameter values from 2.0 mm to 4.0 mm, while the channel heights were fixed between 2.0 mm and 6.0 mm. The optimal ranges of channel height and pinhole diameter were determined by evaluating the intrinsic resolution and sensitivity tradeoff curves. The pinhole parameters were selected based on these analyses, and we verified the simulation results through experimental tests of three types of collimators (general purpose, high sensitivity, and high resolution). The simulated and experimental results agreed, with discrepancies of 4.5% and 6.4% in the sensitivity and spatial resolution, respectively. The results demonstrate that the pinhole collimator designed in this study could be utilized to perform radiation monitoring. -- Highlights: • Optimization of pinhole collimator. • GATE simulation of pinhole gamma camera. • Development of pinhole gamma camera. • Experimental validation

Availability note (English)

Available from http://dx.doi.org/10.1016/j.radmeas.2013.06.004

Additional details

Identifiers

DOI
10.1016/j.radmeas.2013.06.004;
PII
S1350-4487(13)00249-7;

Publishing Information

Journal Title
Radiation Measurements
Journal Volume
59
Journal Page Range
p. 114-118
ISSN
1350-4487
CODEN
RMEAEP

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
45096010
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
BACKGROUND RADIATION; COLLIMATORS; ENVIRONMENT; GAMMA CAMERAS; RADIATION MONITORING; SPATIAL RESOLUTION
Descriptors DEC
CAMERAS; MONITORING; RADIATIONS; RESOLUTION

Optional Information

Copyright
Copyright (c) 2013 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.