A 10% HPGe detector was irradiated with thermal neutrons with a fluence of 4 x 106 n/cm2 to observe neutron-induced gamma-rays in the detector itself; these rays are considered to be the main source of spectral background in neutron in-beam experiments. The significances of the observed peaks in the spectrum were described in terms of both their intensities and detectabilities relative to the 596 keV peak. The contributions of prompt and delayed gamma-rays to the spectral background and the activities remaining in the detector after irradiation are discussed. (author)
Additional details
Publishing Information
- Journal Title
- Applied Radiation and Isotopes
- Journal Volume
- 44
- Journal Issue
- 3
- Journal Page Range
- p. 605-611.
- ISSN
- 0883-2889
- CODEN
- ARISEF
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 24051147
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- GAMMA SPECTRA; HIGH-PURITY GE DETECTORS; IRRADIATION; KEV RANGE 100-1000; NEUTRON FLUENCE; PEAKS; PHYSICAL RADIATION EFFECTS; PROMPT GAMMA RADIATION; THERMAL NEUTRONS
- Descriptors DEC
- BARYONS; ELECTROMAGNETIC RADIATION; ELEMENTARY PARTICLES; ENERGY RANGE; FERMIONS; GAMMA RADIATION; GE SEMICONDUCTOR DETECTORS; HADRONS; IONIZING RADIATIONS; KEV RANGE; MEASURING INSTRUMENTS; NEUTRONS; NUCLEONS; RADIATION DETECTORS; RADIATION EFFECTS; RADIATIONS; SEMICONDUCTOR DETECTORS; SPECTRA