An on-line acoustic fluorocarbon coolant mixture analyzer for the ATLAS silicon tracker
Creators
- 1. Dept. of Physics and Astronomy, Univ. of Glasgow, G12 8QQ (United Kingdom)
- 2. CERN, 1211 Geneva 23 (Switzerland)
- 3. Centre de Physique des Particules de Marseille, 163 Avenue de Luminy, 13288 Marseille Cedex 09 (France)
- 4. Dept. of Physics and Astronomy, Univ. of Oklahoma, Norman, OK 73019 (United States)
- 5. Czech Technical Univ., Technicka 4, 166 07 Prague 6 (Czech Republic)
- 6. Physics Dept., Indiana Univ., Bloomington, IN 47405 (United States)
- 7. B.P. Konstantinov Petersburg Nuclear Physics Inst. PNPI, 188300 St. Petersburg (Russian Federation)
- 8. Rutherford Appelton Laboratory - Science and Technology Facilities Council, Chilton, Didcot OX11 OQX (United Kingdom)
- 9. Graduate School of Pure and Applied Sciences, Univ. of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki 305-8577 (Japan)
Description
The ATLAS silicon tracker community foresees an upgrade from the present octafluoro-propane (C3F8) evaporative cooling fluid - to a composite fluid with a probable 10-20% admixture of hexafluoro-ethane (C2F6). Such a fluid will allow a lower evaporation temperature and will afford the tracker silicon substrates a better safety margin against leakage current-induced thermal runaway caused by cumulative radiation damage as the luminosity profile at the CERN Large Hadron Collider increases. Central to the use of this new fluid is a new custom-developed speed-of-sound instrument for continuous real-time measurement of the C3F8/C2F6 mixture ratio and flow. An acoustic vapour mixture analyzer/flow meter with new custom electronics allowing ultrasonic frequency transmission through gas mixtures has been developed for this application. Synchronous with the emission of an ultrasound 'chirp' from an acoustic transmitter, a fast readout clock (40 MHz) is started. The clock is stopped on receipt of an above threshold sound pulse at the receiver. Sound is alternately transmitted parallel and anti-parallel with the vapour flow for volume flow measurement from transducers that can serve as acoustic transmitters or receivers. In the development version, continuous real-time measurement of C3F8/C2F6 flow and calculation of the mixture ratio is performed within a graphical user interface developed in PVSS-II, the Supervisory, Control and Data Acquisition standard chosen for LHC and its experiments at CERN. The described instrument has numerous potential applications - including refrigerant leak detection, the analysis of hydrocarbons, vapour mixtures for semiconductor manufacture and anesthetic gas mixtures. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1109/ANIMMA.2011.6172830Additional details
Identifiers
Publishing Information
- Publisher
- Inst. of Electrical and Electronics Engineers - IEEE
- Imprint Place
- Piscataway, NJ (United States)
- ISBN
- 978-1-4577-0926-5
- Imprint Pagination
- 7 p.
Conference
- Title
- 2. International Conference on Advancements in Nuclear Instrumentation, Measurement Methods and their Applications
- Acronym
- ANIMMA 2011
- Dates
- 6-9 Jun 2011
- Place
- Ghent (Belgium)
INIS
- Country of Publication
- United States
- Country of Input or Organization
- France
- INIS RN
- 43130167
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CERN LHC; DATA ACQUISITION; ETHANE; EVAPORATIVE COOLING; FLOWMETERS; LEAKAGE CURRENT; LUMINOSITY; MHZ RANGE; MIXTURES; PARTICLE TRACKS; PROPANE; RADIATION EFFECTS; READOUT SYSTEMS; SAFETY MARGINS; SEMICONDUCTOR DEVICES; SEMICONDUCTOR MATERIALS; SOUND WAVES; TRANSDUCERS; VAPORS
- Descriptors DEC
- ACCELERATORS; ALKANES; COOLING; CURRENTS; CYCLIC ACCELERATORS; DISPERSIONS; ELECTRIC CURRENTS; FLUIDS; FREQUENCY RANGE; GASES; HYDROCARBONS; MATERIALS; MEASURING INSTRUMENTS; METERS; OPTICAL PROPERTIES; ORGANIC COMPOUNDS; PHYSICAL PROPERTIES; STORAGE RINGS; SYNCHROTRONS
Optional Information
- Notes
- 9 refs.; IEEE Catalog Number: CFP1124I-CDR