Effect of computed tomography number-relative electron density conversion curve on the calculation of radiotherapy dose and evaluation of Monaco radiotherapy treatment planning system

  • 1. Tehran University of Medical Sciences, Department of Radiotherapy Physics, Cancer Research Centre, Cancer Institute (Iran, Islamic Republic of)
  • 2. Kashan University of Medical Sciences, Department of Medical Physics and Radiology, Faculty of Paramedical Sciences (Iran, Islamic Republic of)
  • 3. Shahid Beheshti University of Medical Sciences, Cancer Research Center (Iran, Islamic Republic of)
  • 4. Qom University of Medical Sciences, Department of Radiotherapy Physics, Cancer Institute (Iran, Islamic Republic of)
  • 5. Comprehensive Cancer Centers of Nevada (United States)

Description

The accuracy of a computed tomography (CT)-relative electron density (RED) curve may have an indirect impact on the accuracy of dose calculation by a treatment planning system (TPS). This effect has not been previously quantified for input of different CT-RED curves from different CT-scan units in the Monaco TPS. This study aims to evaluate the effect of CT-RED curve on the dose calculation by the Monaco radiotherapy TPS. Four CT images of the CIRS phantom were obtained by different CT scanners. The accuracy of the dose calculation in the three algorithms of the Monaco TPS (Monte Carlo, collapse cone, and pencil beam) is also evaluated based on TECDOC 1583. The CT-RED curves from the CT scanners were transferred to the Monaco TPS to audit the different algorithms of the TPS. The dose values were measured with an ionization chamber in the CIRS phantom. Then, the dose values were calculated by the Monaco algorithms in the corresponding points. For the Monaco TPS and based on TECDOC 1583, the accuracy of the dose calculation in all the three algorithms is within the agreement criteria for most of the points evaluated. For low dose regions, the differences between the calculated and measured dose values are higher than the agreement criteria in a number of points. For the majority of the points, the algorithms underestimate the calculated dose values. It was also found that the use of different CT-RED curves can lead to minor discrepancies in the dose calculation by the Monaco TPS, especially in low dose regions. However, it appears that these differences are not clinically significant in most of the cases.

Additional details

Identifiers

Publishing Information

Journal Title
Australasian Physical and Engineering Sciences in Medicine
Journal Volume
42
Journal Issue
2
Journal Page Range
p. 489-502
ISSN
0158-9938
CODEN
AUPMDI

Optional Information

Copyright
Copyright (c) 2019 Australasian College of Physical Scientists and Engineers in Medicine