Published September 2006 | Version v1
Miscellaneous Open

Establishment of simulated mammography radiation qualities at the BIPM

Creators

  • 1. Bureau International des Poids et Mesures (BIPM), 92 - Sevres (France)

Description

New radiation qualities have been established at the BIPM using the tungsten-anode low-energy x-ray tube with molybdenum and rhodium as filters to simulate the radiation beams used in clinical mammography. The evaluation of the correction factors for the primary standard and the measurements of the x-ray spectra are described. (authors)

Availability note (English)

Available from INIS in electronic form

Files

38101991.pdf

Files (329.9 kB)

Name Size Download all
md5:6c586dedd0261e365e6593d4ef92166b
329.9 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
8 p.
Report number
BIPM--06-08

Optional Information

Notes
5 refs.