Published 1980 | Version v1
Report Open

Calculation of leading-edge pulse shaper time resolution for detectors with internal amplification

Description

The problem of time data readout from scintillation and Cherenkov detectors, from proportional and drift chambers has completely been solved taking account of shaper time slowing-down. The real amplitude distribution and the shape of current signals from detectors are considered. The results of calculation are applicable for finding time resolution, when current pulses are integrated by capacity of detector, or long cable, which connects a detector with pulse shaper

Availability note (English)

MF available from INIS under the Report Number.

Files

12575889.pdf

Files (280.9 kB)

Name Size Download all
md5:1436c35806785f45c7b3d6c61b35bc68
280.9 kB Preview Download

Additional details

Additional titles

Original title (Russian)
Raschet vremennogo razresheniya formirovatelej po perednemu frontu impul'sov detektorov s vnutrennim usileniem

Publishing Information

Imprint Pagination
12 p.
Report number
JINR--13-80-263

Optional Information

Notes
11 refs.; 5 figs.; submitted to the journal Instrum. Exp. Tech.