Published November 2017 | Version v1
Journal article

Development of continuous inflow tritium measurement in water technology using electrolysis and a plastic scintillator

  • 1. Ulsan National Institute of Science and Technology, Ulsan (Korea, Republic of). Department of Nuclear Engineering

Description

Our goal was to develop a mobile tritium monitor for continuous inflow system for water sample. The system is based on electrolysis and a plastic scintillator detection system. The minimum detectable activity (MDA) of the prototype system is 431 kBq L-1, while the MDA of a commercially available product is 740 kBq L-1. We expected to achieve a 5.73-times lower MDA by optimizing detection geometry using a multi-hydrogen-gas-channel. The system can be applied either as a mobile leakage surveying method or as a fixed-type monitor for detecting tritium in drinking water by adapting conventional background reduction technologies. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
314
Journal Issue
2
Journal Page Range
p. 689-694
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
23 refs.; This record replaces 48095793