Published May 1988
| Version v1
Journal article
Event selection unit according to pulse shape of proportional counter
Description
An event selection unit according to pulse shape of a proportional counter with xenon filling is proposed. The use of a peak detector has enabled to realize direct measurement of pulse front duration. Background count rate suppression in the 2-12 keV energy range is 89 %, when the loss of efficiency of useful signal detection (which is simulated by 5.9 keV γ-quanta) is 11 %
Additional details
Additional titles
- Original title (Russian)
- Устройство отбора событий по форме импульсов пропорционального счетчика
Publishing Information
- Journal Title
- Pribory i Tekhnika Ehksperimenta
- Journal Issue
- no.3
- Series
- Prib. Tekh. Ehksp.
- ISSN
- 0032-8162
- CODEN
- PRTEA
INIS
- Country of Publication
- Russian Federation
- Country of Input or Organization
- USSR
- INIS RN
- 20023582
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BACKGROUND NOISE; EFFICIENCY; ENERGY RESOLUTION; FLOWSHEETS; MULTIWIRE PROPORTIONAL CHAMBER; PULSE ANALYZERS; PULSE RISE TIME; SOFT X RADIATION; X-RAY DETECTION; XENON
- Descriptors DEC