The influence of crosstalk on the fast fluid attenuated inversion recovery (fast FLAIR) pulse sequence was studied. On multislice fast FLAIR images, the water null point was shortened in comparison with that on single slice images owing to the crosstalk received from adjacent slices. That influence became greater with decreases in the slice gap and increases in the number of slices. The timing of crosstalk in each slice varied according to excitation order. The process of recovery of longitudinal magnetization changed according to differences in timing; thus, it was possible that the water null point changed in each slice. In brain images with thinner slice gaps, the signal intensity of CSF is increased by the effect of crosstalk. In order to eliminate changes in the water null point caused by crosstalk on fast FLAIR, the multislice sequence requires a sequence with interleaving based on the premise that slice gaps are set for more than 100% of slice thickness. (author)