Published December 2014 | Version v1
Journal article

Implementation of GaN based real-time source position monitoring in HDR brachytherapy

  • 1. Université Lyon 1, Villeurbanne F-69622 (France)
  • 2. Institut des Nanotechnologies de Lyon, INL, CNRS UMR5270, F-69003 (France)
  • 3. Centre Hospitalier Lyon Sud, BP 217 Lyon (France)

Description

For High Dose Rate (HDR) brachytherapy Quality Assurance (QA), we propose a seed position verification method by using two GaN dosimetric probes integrated in an instrumented phantom. As dose rate measured from one probe is related to the seed distance, seed position inside catheter can be determined. The proposed two-probe method extends measureable range and improves accuracy in comparison with the use of one single probe. One instrumented phantom implementing the method was designed, fabricated and tested in clinical conditions. The measureable range of the two-probe method is 105 mm compared to 36 mm using single probe. The validity of the method was experimentally verified. - Highlights: • A simple brachytherapy QA phantom has been instrumented. • Dwell positions have been estimated according to measured dose rate by GaN probes. • Measured dose rate shows 1% average deviation from calculated results. • Determined dwell position precision and measureable range was studied

Availability note (English)

Available from http://dx.doi.org/10.1016/j.radmeas.2014.05.021

Additional details

Identifiers

DOI
10.1016/j.radmeas.2014.05.021;
PII
S1350-4487(14)00159-0;

Publishing Information

Journal Title
Radiation Measurements
Journal Volume
71
Journal Page Range
p. 293-296
ISSN
1350-4487
CODEN
RMEAEP

Conference

Title
17. solid state dosimetry conference
Acronym
SSD17
Dates
22-27 Sep 2013
Place
Recife (Brazil)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
46127957
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ACCURACY; BRACHYTHERAPY; COMPARATIVE EVALUATIONS; DOSE RATES; GALLIUM NITRIDES; MONITORING; PHANTOMS; QUALITY ASSURANCE; RADIATION SOURCE IMPLANTS; VERIFICATION
Descriptors DEC
EVALUATION; GALLIUM COMPOUNDS; IMPLANTS; MEDICINE; MOCKUP; NITRIDES; NITROGEN COMPOUNDS; NUCLEAR MEDICINE; PNICTIDES; RADIATION SOURCES; RADIOLOGY; RADIOTHERAPY; STRUCTURAL MODELS; THERAPY

Optional Information

Copyright