Published March 1980
| Version v1
Journal article
Evaluation of edge-induced streaking artifacts in CT scanners
- 1. Neuroradiology and Computed Tomography Section, Building 10, Room 11N242, National Institute of Health, Bethesda, Maryland 20205
Description
A phantom is presented which permits the evaluation of streaking artifacts produced in CT reconstructions by abrupt edges. Its application is demonstrated by results obtained from nine CT scanners. It is observed that even in regions where streaking is not readily apparent, edge-induced artifacts can significantly increase the variance in the reconstruction
Additional details
Publishing Information
- Journal Title
- Med. Phys.
- Journal Volume
- 7
- Journal Issue
- 2
- Series
- Med. Phys.
- Journal Page Range
- 108-111
- ISSN
- 0094-2405
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 11541343
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- IMAGE PROCESSING; PHANTOMS; TOMOGRAPHY
- Descriptors DEC