Experimental measurement of radiation dose in a dedicated breast CT system
Creators
- 1. Beijing Engineering Research Center of Radiographic Techniques and Equipment (China)
- 2. Key Laboratory of Nuclear Analytical Techniques, Institute of High Energy Physics, Chinese Academy of Sciences, Beijing (China)
- 3. College of Nuclear Science and Technology, Harbin Engineering University (China)
Description
Radiation dose is an important performance indicator of a dedicated breast CT (DBCT). In this paper, the method of putting thermoluminescent dosimeters (TLD) into a breast shaped PMMA phantom to study the dose distribution in breasts was improved by using smaller TLDs and a new half-ellipsoid PMMA phantom. Then the weighted CT dose index (CTDIw) was introduced to average glandular assessment in DBCT for the first time and two measurement modes were proposed for different sizes of breasts. The dose deviations caused by using cylindrical phantoms were simulated using the Monte Carlo method and a set of correction factors were calculated. The results of the confirmatory measurement with a cylindrical phantom (11 cm/8 cm) show that CTDIw gives a relatively conservative overestimate of the average glandular dose comparing to the results of Monte Carlo simulation and TLDs measurement. But with better practicability and stability, the CTDIw is suitable for dose evaluations in daily clinical practice. Both of the TLDs and CTDIw measurements demonstrate that the radiation dose of our DBCT system is lower than conventional two-view mammography. (authors)
Availability note (English)
Available from DOI: http://dx.doi.org/10.1088/1674-1137/38/3/038201Additional details
Identifiers
Publishing Information
- Journal Title
- Chinese Physics. C, High Energy Physics and Nuclear Physics
- Journal Volume
- 38
- Journal Issue
- 3
- Journal Page Range
- [7 p.]
- ISSN
- 1674-1137
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 47036035
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; CAT SCANNING; COMPARATIVE EVALUATIONS; COMPUTERIZED SIMULATION; CORRECTIONS; CYLINDRICAL CONFIGURATION; MAMMARY GLANDS; MONTE CARLO METHOD; PERFORMANCE; PHANTOMS; PMMA; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; STABILITY; THERMOLUMINESCENT DOSEMETERS; WEIGHT
- Descriptors DEC
- BODY; CALCULATION METHODS; COMPUTERIZED TOMOGRAPHY; CONFIGURATION; DIAGNOSTIC TECHNIQUES; DOSEMETERS; DOSES; ESTERS; EVALUATION; GLANDS; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; ORGANIC COMPOUNDS; ORGANIC POLYMERS; ORGANS; POLYACRYLATES; POLYMERS; POLYVINYLS; RADIOLOGY; SIMULATION; STRUCTURAL MODELS; TOMOGRAPHY
Optional Information
- Notes
- 4 figs., 4 tabs., 13 refs.; This record replaces 46127762