Published April 1, 2002 | Version v1
Journal article

Wien filter for cooled low-energy radioactive ion beams

Description

A Wien filter for cooled radioactive ion beams has been designed at Ion Guide Isotope Separator On Line technique (IGISOL). The purpose of such device is to eliminate doubly charged ions from the mass separated singly charged ions, based on q=+2→q=+1 charge exchange process in an ion cooler. The performance of the Wien filter has been tested off-line with a discharge ion source as well as on-line with a radioactive beam. The electron capture process of cooled q=+2 ions has been investigated in a radiofrequency quadrupole ion cooler with varying partial pressures of nitrogen. Also, the superasymmetric fission production yields of 68<A<78 nuclei have been deduced

Additional details

Identifiers

PII
S0168900201013626;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
481
Journal Issue
1-3
Journal Page Range
p. 718-730
ISSN
0168-9002
CODEN
NIMAER

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
34003489
Subject category
S07: ISOTOPES AND RADIATION SOURCES;
Descriptors DEI
ATOMIC NUMBER; ELECTRON CAPTURE; FILTERS; FISSION PRODUCTS; HEAT EXCHANGERS; MULTICHARGED IONS; NITROGEN; RADIOACTIVE ION BEAMS
Descriptors DEC
BEAMS; CAPTURE; CHARGED PARTICLES; ELEMENTS; ION BEAMS; IONS; ISOTOPES; MATERIALS; NONMETALS; RADIOACTIVE MATERIALS

Optional Information

Copyright
Copyright (c) 2002 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.