Method on determination of wall-thickness of two-dimensional ionization chamber dose detector based on EGSnrc
Creators
- 1. Institute of Nuclear and New Energy Technology, Tsinghua University, Beijing (China)
- 2. Beijing Key Laboratory of Nuclear Detection and Measurement Technology, Beijing (China)
Description
The wall-thickness is an important design parameter for two-dimensional ionization chamber dose detector. Under the given geometric parameters of the cavity unit of two-dimensional ionization chamber dose detector, the simulation model was built by using Monte Carlo method. The wall-thickness was calculated with the simulation model using EGSnrc, and the reference numerical value was given. The result shows that when the cavity is a cylinder with diameter of 4.5 mm and height of 5 mm, under radiation of 1.25 MeV, the front wall-thickness of ionization chamber should be more than 4.4 mm and the side wall-thickness should be 1.5 mm in order to meet the demand of accuracy of 1%. Under the radiation with energy of 1.25-4 MeV, the side wall-thickness should be more than 5 mm in order to meet the demand of accuracy of 1%. The research result can be used as the design basis for the detector. (authors)
Additional details
Identifiers
Publishing Information
- Journal Title
- Atomic Energy Science and Technology
- Journal Volume
- 52
- Journal Issue
- 4
- Journal Page Range
- p. 710-714
- ISSN
- 1000-6931
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 55027356
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACCURACY; CAVITIES; CYLINDERS; DOSES; GEOMETRY; HEIGHT; IONIZATION CHAMBERS; MEV RANGE; MONTE CARLO METHOD; THICKNESS; TWO-DIMENSIONAL CALCULATIONS; WALLS
- Descriptors DEC
- CALCULATION METHODS; DIMENSIONS; ENERGY RANGE; MATHEMATICS; MEASURING INSTRUMENTS; RADIATION DETECTORS
Optional Information
- Notes
- 6 figs., 3 tabs., 10 refs.; http://dx.doi.org/10.7538/yzk.2017.youxian.0473