There is a newer version of the record available.

Published April 2018 | Version v1
Journal article

Method on determination of wall-thickness of two-dimensional ionization chamber dose detector based on EGSnrc

  • 1. Institute of Nuclear and New Energy Technology, Tsinghua University, Beijing (China)
  • 2. Beijing Key Laboratory of Nuclear Detection and Measurement Technology, Beijing (China)

Description

The wall-thickness is an important design parameter for two-dimensional ionization chamber dose detector. Under the given geometric parameters of the cavity unit of two-dimensional ionization chamber dose detector, the simulation model was built by using Monte Carlo method. The wall-thickness was calculated with the simulation model using EGSnrc, and the reference numerical value was given. The result shows that when the cavity is a cylinder with diameter of 4.5 mm and height of 5 mm, under radiation of 1.25 MeV, the front wall-thickness of ionization chamber should be more than 4.4 mm and the side wall-thickness should be 1.5 mm in order to meet the demand of accuracy of 1%. Under the radiation with energy of 1.25-4 MeV, the side wall-thickness should be more than 5 mm in order to meet the demand of accuracy of 1%. The research result can be used as the design basis for the detector. (authors)

Additional details

Publishing Information

Journal Title
Atomic Energy Science and Technology
Journal Volume
52
Journal Issue
4
Journal Page Range
p. 710-714
ISSN
1000-6931

Optional Information

Notes
6 figs., 3 tabs., 10 refs.; http://dx.doi.org/10.7538/yzk.2017.youxian.0473