Published August 2018 | Version v1
Journal article

Neutron sensitivity of 10B4C-coated aluminum honeycomb using a single-anode wire, P-10 continuous-gas-flow proportional counter

  • 1. S.M.A.R.T. Laboratory, Department of Mechanical and Nuclear Engineering, Kansas State University, Manhattan, KS 66505 (United States)

Description

The intrinsic thermal-neutron detection efficiency of a 10B4C-coated aluminum honeycomb sample was measured using a single-anode wire, P-10 continuous-gas-flow proportional counter. The aluminum honeycomb sample was 1.27 cm thick with a diameter of 4.1275 cm (1 5/8 in.) and was composed of hexagonal-shaped cells with a cell size of 1.5875 mm (1/16 in.). The 10B4C coating was applied by Exothermics, Inc. via magnetron sputtering to a coating thickness of 4.68±1.25μm as measured using a scanning electron microscope. Using a lower level discriminator setting of 50 keV, intrinsic thermal-neutron detection efficiency was measured to be 21.45 ± 0.26% in reference to a 4.0 atm, 5.08-cm (2-in.) diameter, 15.24-cm (6-in.) long Reuter Stokes 3He tube with a known efficiency of 80.7 ± 0.5%. MCNP6 simulations yielded a theoretical intrinsic thermal-neutron detection efficiency of approximately 24% for a 10B4C-coating thickness of 4.68μm with a lower level discriminator setting of 50 keV.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2018.04.046

Additional details

Identifiers

DOI
10.1016/j.nima.2018.04.046;
PII
S0168900218305540;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
898
Journal Page Range
p. 85-89
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2018 Elsevier B.V. All rights reserved.