Published August 2012 | Version v1
Book

Determination the Content of Palladium of Pd-Al2O3 Pellets by EDXRF

  • 1. China Academy of Engineering Physics, Mianyang (China)

Description

A simple, fast, and accurate method of determination the content of palladium of Pd-Al2O3 pellets by EDXRF was established. By added the internal standard element Mo quantitatively to the sample, dissolved Pd on the Pd-Al2O3 pellets using aqua regia, and made filter paper sample, the content of Pd could be obtained Energy dispersive X-Ray Fluorescence Spectrometer. While the content of palladium between 30 wt% - 50 wt% of the sample, the relative standard deviation less than 0.9% and the relevance ratio between 98%-102%. The measuring range of this method can be adjusted by changing the content of Mo. (authors)

Part of:
Proceedings of 16. national conference on nuclear electronics and nuclear detection technology (part 2)

Additional details

Publishing Information

Publisher
Chinese Nuclear Society
Imprint Place
Beijing (China)
Imprint Title
Proceedings of 16. national conference on nuclear electronics and nuclear detection technology (part 2)
Imprint Pagination
299 p.
Journal Page Range
p. 517-520

Conference

Title
16. national conference on nuclear electronics and nuclear detection technology
Dates
15-16 Aug 2012
Place
Mianyang (China)

Optional Information

Notes
2 figs., 3 tabs., 8 refs.