Published May 1987 | Version v1
Journal article

Position-sensitive neutron detectors based on a scintillator and a device with charge coupling

  • 1. Leningrad Institute of Nuclear Physics, USSR

Description

A position-sensitive neutron detector on the basis of a scintillator and a device with charge coupling was developed. A neutron flux with an intensity of 105 cm-2 sec-1 was recorded. The resolution is 0.4 mm. The size of the sensitive region, which in the present setup is 3.9 x 4.8 mm, can be increased with the aid of a focon or with a high-power lens

Additional details

Publishing Information

Journal Title
Instrum. Exp. Tech. (Engl. Transl.)
Journal Volume
29
Journal Issue
6
Series
Instrum. Exp. Tech. (Engl. Transl.).
Journal Page Range
1297-1299
ISSN
0020-4412
CODEN
INETA

Optional Information

Notes
Translated from Prib. Tekh. Eksp.; 29: No. 6, 56-57(Nov-Dec 1986).