Published December 2014 | Version v1
Journal article

Radiation dose associated with cerebral CT angiography and CT perfusion: an experimental phantom study

  • 1. Diagnostic Imaging and Radiotherapy Program, School of Diagnostic and Applied Health Sciences, Faculty of Health Sciences, Universiti Kebangsaan Malaysia, Kuala Lumpur 50300 (Malaysia)
  • 2. Discipline of Medical Imaging, Department of Imaging and Applied Physics, Curtin University, GPO Box U1987 Perth, WA 6845 (Australia)

Description

A study on the radiation dose associated with cerebral CT angiography (CTA) and CT perfusion (CTP) was conducted on an anthropomorphic phantom with the aim of estimating the effective dose (E) and entrance skin dose (ESD) in the eyes and thyroid gland during different CTA and CTP protocols. The E was calculated to be 0.61 and 0.28 mSv in CTA with 100 and 80 kVp, respectively. In contrast, CTP resulted in an estimated E of 2.74 and 2.07 mSv corresponding to 40 and 30 s protocols, respectively. The eyes received a higher ESD than the thyroid gland in all of these protocols. The results of this study indicate that combining both CTA and CTP procedures are not recommended in the stroke evaluation due to high radiation dose. Application of modified techniques in CTA (80 kVp) and CTP (30 s) is highly recommended in clinical practice for further radiation dose reduction. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/nct280

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
162
Journal Issue
3
Journal Page Range
p. 316-321
ISSN
0144-8420

Optional Information

Notes
22 refs.