Published January 2012 | Version v1
Journal article

Case study thoracic radiotherapy in an elderly patient with pacemaker: The issue of pacing leads

  • 1. Department of Radiation Oncology, Institut Curie, Paris (France)
  • 2. Department of Anesthesiology and Reanimation, Institut Curie, Paris (France)

Description

To assess clinical outcome of patients with pacemaker treated with thoracic radiation therapy for T8-T9 paravertebral chloroma. A 92-year-old male patient with chloroma presenting as paravertebral painful and compressive (T8-T9) mass was referred for radiotherapy in the Department of Radiation Oncology, Institut Curie. The patient presented with cardiac dysfunction and a permanent pacemaker that had been implanted prior. The decision of Multidisciplinary Meeting was to deliver 30 Gy in 10 fractions for reducing the symptoms and controlling the tumor growth. The patient received a total dose of 30 Gy in 10 fractions using 4-field conformal radiotherapy with 20-MV photons. The dose to pacemaker was 0.1 Gy but a part of the pacing leads was in the irradiation fields. The patient was treated the first time in the presence of his radiation oncologist and an intensive care unit doctor. Moreover, the function of his pacemaker was monitored during the entire radiotherapy course. No change in pacemaker function was observed during any of the radiotherapy fractions. The radiotherapy was very well tolerated without any side effects. The function of the pacemaker was checked before and after the radiotherapy treatment by the cardiologist and no pacemaker dysfunction was observed. Although updated guidelines are needed with acceptable dose criteria for implantable cardiac devices, it is possible to treat patients with these devices and parts encroaching on the radiation field. This case report shows we were able to safely treat our patient through a multidisciplinary approach, monitoring the patient during each step of the treatment.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.meddos.2011.07.001

Additional details

Identifiers

DOI
10.1016/j.meddos.2011.07.001;
PII
S0958-3947(11)00150-6;

Publishing Information

Journal Title
Medical Dosimetry
Journal Volume
37
Journal Issue
2
Journal Page Range
p. 192-194
ISSN
0958-3947

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
44085527
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
CARDIAC PACEMAKERS; DOSIMETRY; IRRADIATION; LEAD; MEDICAL PERSONNEL; NEOPLASMS; PATIENTS; PHOTONS; RADIATION DOSES; RADIOTHERAPY; RECOMMENDATIONS; SIDE EFFECTS; SYMPTOMS
Descriptors DEC
BOSONS; DISEASES; DOSES; ELEMENTARY PARTICLES; ELEMENTS; MASSLESS PARTICLES; MEDICINE; METALS; NUCLEAR MEDICINE; PERSONNEL; RADIOLOGY; THERAPY

Optional Information

Copyright
Copyright (c) 2012 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.