Published October 1990 | Version v1
Journal article

Real-time digital control, data acquisition, and analysis system for the DIII-D multipulse Thomson scattering diagnostic

  • 1. General Atomics, San Diego, CA (USA)

Description

A VME-based real-time computer system for laser control, data acquisition, and analysis for the DIII-D multipulse Thomson scattering diagnostic is described. The laser control task requires precise timing of up to eight Nd:YAG lasers, each with an average firing rate of 20 Hz. A cpu module in a real-time multiprocessing computer system will operate the lasers with evenly staggered laser pulses or in a ''burst mode,'' where all available (fully charged) lasers can be fired at 50--100 μs intervals upon receipt of an external event trigger signal. One or more cpu modules, along with a LeCroy FERA (fast encoding and readout ADC) system, will perform real-time data acquisition and analysis. Partial electron temperature and density profiles will be available for plasma feedback control within 1 ms following each laser pulse. The VME-based computer system consists of two or more target processor modules (25 MHz Motorola 68030) running the VMEexec real-time operating system connected to a Unix-based host system (also a 68030). All real-time software is fully interrupt driven to maximize system efficiency. Operator interaction and (non-real-time) data analysis takes place on a MicroVAX 3400 connected via DECnet

Additional details

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
61
Journal Issue
10
Series
Rev. Sci. Instrum.
Journal Page Range
3286-3288
ISSN
0034-6748
CODEN
RSINA