There is a newer version of the record available.

Published October 2023 | Version v1
Journal article

Simulation of Hp(0,07) beta and photon response of a BeOSL finger ring dosemeter

  • 1. Mirion Medical GmbH, Munich 81739 (Germany)

Description

Simulation of a dosemeters energy and angular response lowers development cost and time. To support development of a new beta sensitive dosemeter we simulate both the dosemeter and the skin dose in a rod phantom as there are no fluence to dose conversion coefficients in the ICRU Report 47 for electrons. Published spectra for BSS2, N-Series and S-Cs and S-Co were used to define sources, with CAD models used to define the simulation geometry. Simulation tools were validated with experimental beta and photon responses available for the BeOSL photon ring. The developed tool chain was then used to simulate several iterations of a beta sensitive detector element inside a beta ring holder, with the goal to minimise the need for repeated prototype fabrication and characterisation. Simulation results for the final design were compared to measurements on 3D printed prototypes with all radiation qualities, except 90Sr/90Y 0 deg., falling within measurement uncertainty. (authors)

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
199
Journal Issue
15-16
Journal Page Range
p. 1770-1773
ISSN
0144-8420

Conference

Title
Joint International Conference on Individual Monitoring of Ionizing Radiation; NEUDOS-14: Neutron and Ion Dosimetry Symposium
Acronym
IM2022
Dates
25-29 Apr 2022
Place
Krakow (Poland)

Optional Information

Notes
15 refs.