Published February 2017 | Version v1
Journal article

The elementary analysis on several problems of medical accelerator charging auxiliary circuit design

  • 1. Nuclear power of China Equipment manufacturing Factory. Sichuan Xihe Nuclear Mechanical & Electrical Equipment Manufacturing Co., LTD. Chengdu (China)

Description

Objective: To find out the principle and method of the design of charging auxiliary circuit of medical accelerators. Methods: The method is mainly base on the physical principle and trying to avoid the complex mathematical formula in the analysis of waveform frequency and waveform source circuit. Results: Through the use of critical damping circuit, complete the full adjustment of pulse modulator. Conclusion: Through the analysis of the main factor of the circuit, the actual auxiliary circuit design and verification are consistent and obtained a good verification. (authors)

Additional details

Publishing Information

Journal Title
Chinese Journal of Radiological Health
Journal Volume
26
Journal Issue
1
Journal Page Range
p. 96-99
ISSN
1004-714X

INIS

Country of Publication
China
Country of Input or Organization
China
INIS RN
50069815
Subject category
S43: PARTICLE ACCELERATORS;
Descriptors DEI
ACCELERATORS; DESIGN; ELECTRONIC CIRCUITS; MEDICINE; PULSES; VERIFICATION

Optional Information

Notes
13 figs., 3 refs.