Microsecond and nanosecond pulse electorn beam parameters measuring

  • 1. Leningradskij Tekhnologicheskij Inst. (USSR)

Description

Investigation results of dosimetric properties of the TATs-140 industrial film on the base of cellulose triacetate are presented. Film irradiation has been conducted in ''Elit-2'' and ''Minus'' accelerators with 1.5 MeV and 400 keV electron beam 12.5 A and 5 kA been current, 3 μs and 10 ns current pulse duration respectively. It is shown that the TATs-140 film can be used as a dosimetric system for electron pulsed beams of microsecond range with 5x109 radxs-1 dose rate. Linearity of the absorbed dose is preserved from 3 to 100 mrad with +-10% error

Additional details

Additional titles

Original title (Russian)
Измерение параметров импульсных электронных пучков микросекундного и наносекундного диапазонов

Publishing Information

Imprint Title
Reports of 3. All-union conference on application of charged particles accelerators in national economy. V. 3. 3
Journal Page Range
p. 263-267.
Report number
INIS-SU--74

Conference

Title
3. All-union conference on application of charged particles accelerators in national economy. V. 3.
Dates
26 - 28 Jun 1979.
Place
Leningrad, USSR.

INIS

Country of Publication
USSR
Country of Input or Organization
USSR
INIS RN
13665420
Subject category
S43: PARTICLE ACCELERATORS;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ACCURACY; AMP BEAM CURRENTS; CELLULOSE; DOSE RATES; DOSEMETERS; ELECTRON BEAMS; FILMS; KEV RANGE 100-1000; KILO AMP BEAM CURRENTS; RADIATION DOSES; SPACE DEPENDENCE
Descriptors DEC
BEAM CURRENTS; BEAMS; CARBOHYDRATES; CURRENTS; ENERGY RANGE; KEV RANGE; LEPTON BEAMS; MEASURING INSTRUMENTS; ORGANIC COMPOUNDS; PARTICLE BEAMS; POLYSACCHARIDES; SACCHARIDES

Optional Information

Notes
3 refs.; 2 figs. Imprint:Doklady 3. Vsesoyuznogo soveshchaniya po primeneniyu uskoritelej zaryazhennykh chastits v narodnom khozyajstve. Tom 3.