TU-B-304-00: The Aftermath of TG-142
Creators
Description
Although published in 2009, the AAPM TG-142 report on accelerator quality assurance still proves a challenge for full clinical implementation. The choice of methodologies to satisfy TG-142 requirements is critical to a successful application. Understanding the philosophy of TG-142 can help in creating an institution-specific QA practice that is both efficient and effective. The concept of maintaining commissioned beam profiles is still found confusing. The physicist must also consider technologies not covered by TG-142 (i.e. arc therapy techniques). On the horizon is TG-198 report on implementing TG-142. Although the community still lacks a final TG-100 report, performing a failure-mode -and-effects analysis and statistical process control analysis to determine the institution-specific clinical impact of each TG-142 test may be useful for identifying trends for pro-active surveillance. Learning Objectives: To better understand the confusing and controversial aspects of TG-142. To understand what is still missing from TG-142 and how to account for these tests in clinical practice To describe which QA tests in TG-142 yield the largest potential clinical result if not discovered
Additional details
Identifiers
- DOI
- 10.1118/1.4925561;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 42
- Journal Issue
- 6
- Journal Page Range
- vp.
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 48007177
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S60: APPLIED LIFE SCIENCES;
- Descriptors DEI
- ACCELERATORS; BEAM PROFILES; FAILURES; INSPECTION; LEARNING; MEDICAL SURVEILLANCE; PROCESS CONTROL; QUALITY ASSURANCE; RADIOTHERAPY
- Descriptors DEC
- CONTROL; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY
Optional Information
- Notes
- (c) 2015 American Association of Physicists in Medicine