Published October 2015 | Version v1
Miscellaneous

A Field Test of the New Portable Gamma Spectrometry System

  • 1. Korea Institute of Nuclear Non-proliferation and Control, Daejeon (Korea, Republic of)
  • 2. NuCare Medical Systems, Inc., Incheon (Korea, Republic of)

Description

In order to perform a field test of the system, the measurement of U-235 enrichment for nuclear fuel pellets was conducted along with the IAEA Physical Inventory Verification (PIV) inspection at the KEPCO Nuclear Fuel (KNF). The enrichment value of U-235 was calculated based on the total counts of the 185.7 keV photopeak and compared with the reference line, drawn by certified sources. The goal of this study is to experimentally evaluate the system performance of the developed system. In this study, the new portable gamma spectrometry system showed a good linearity (R2=1) but overestimated the enrichment values than IAEA inspection device. It could be caused by the stability of the new system since it found, right after this measurement, that the accuracy of the system gradually increases and becomes stable over time. Further steps will optimize the design parameter based on these results and repeat measurement with the same samples under the same environment

Part of:
Proceedings of the KNS 2015 Fall Meeting

Additional details

Publishing Information

Publisher
KNS
Imprint Place
Daejeon (Korea, Republic of)
Imprint Title
Proceedings of the KNS 2015 Fall Meeting
Imprint Pagination
[1 CD-ROM]
Journal Page Range
[2 p.]

Conference

Title
2015 Fall meeting of the KNS
Dates
28-30 Oct 2015
Place
Kyungju (Korea, Republic of)

Optional Information

Notes
3 refs, 4 figs, 2 tabs