Published 2011 | Version v1
Book

A novel readout module for single photon solid state detectors (SiPMD, GAPD, MPPC, MAPC)

  • 1. Nuclear Physics Inst. of Academy Science of Czech Republic, Rez, Prague 25068 (Czech Republic)
  • 2. Czech Technical Univ., Prague (Czech Republic)

Description

In this paper a novel, Readout Module (RM) for Single Photon Detector (SiPD has been described. The electronics design is based on the concept of virtual instrumentation RM consists of SiPD preamplifier, shaping amplifier, discriminator, multi channel analyzer and control module connected to a PC through the USB bus and of PC application software. The RM can be used for investigation of different types of SiPD with maximum biasing voltage 90 V and maximal current 2 mA. The RM has fast digital output for triggering and 12 bit internal ADC for output digitizing. The RM uses USB bus as a power supply. It could be very useful for laboratory experiment. The small size of module allows easy integration of few modules into multi-channel system that can be used for PET application. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1109/ANIMMA.2011.6172962

Additional details

Identifiers

Publishing Information

Publisher
Inst. of Electrical and Electronics Engineers - IEEE
Imprint Place
Piscataway, NJ (United States)
ISBN
978-1-4577-0926-5
Imprint Pagination
2 p.

Conference

Title
2. International Conference on Advancements in Nuclear Instrumentation, Measurement Methods and their Applications
Acronym
ANIMMA 2011
Dates
6-9 Jun 2011
Place
Ghent (Belgium)

Optional Information

Notes
3 refs.; IEEE Catalog Number: CFP1124I-CDR