Published January 1, 2018 | Version v1
Journal article

Determination of glucose in human urine by cyclic voltammetry method using gold electrode

  • 1. Department of Chemistry, Faculty of Mathematics and Natural Sciences, Universitas Islam Indonesia, Jalan Kaliurang KM 14,5 Sleman Yogyakarta, Indonesia. (Indonesia)

Description

This study has been the determination of glucose in human urine by cyclic voltammetry method using gold electrode. The gold electrode was prepared using gold wire with purity 99.99%, size 1.0 mm by length and wide respectively, connected with silver wire using silver conductive paint. The effect of electrolyte, pH and glucose concentration has been determined to produce the optimum method. The research showed the KNO3 is a good electrolyte for determination of glucose in human urine using gold electrode. The effect of glucose concentration have the coefficient correlation is R2 = 0.994. The results of the recovery using addition method showed at range95-105%. As a conclusion isa gold electrode is a good electrode for electrochemical sensors to the determination of glucose in human urine. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1757-899X/299/1/012001

Additional details

Publishing Information

Journal Title
IOP Conference Series. Materials Science and Engineering (Online)
Journal Volume
299
Journal Issue
1
Journal Page Range
[7 p.]
ISSN
1757-899X

Conference

Title
International Conference on Chemistry and Material Science
Acronym
IC2MS 2017
Dates
4-5 Nov 2017
Place
Malang (Indonesia)