Published 1990 | Version v1
Book

Noise equivalent count measurements in a neuro-PET scanner with retractable septa

  • 1. MCR Cyclotron Unit, Hammersmith Hospital, London (UK)
  • 2. CNR, Inst. Technologie Biomediche Avanzat Medicina Nucleare, Milano (Italy)
  • 3. Dept. of Nuclear Medicine, Univ. Hospital of Geneva, Geneva (Switzerland)

Description

This paper reports on the removal of interplane septa in a PET scanner that enables acquisition of all possible lines of response (3D mode) in an effort to maximize the available number of detected events. One problem with this method at high countrates, however, is a markedly increased deadtime and randoms rate, which has a deleterious effect on data quality. The noise-equivalent countrate (NEC) performance of a neuro-PET scanner has been determined with and without interplane septa on uniform cylindrical phantoms of differing radii and in human studies to assess the optimum countrate conditions that realize the maximum gain. In the brain, the effective gain in NEC performance for 3D ranges from >5 at low countrates to ∼3.3 at 200 kcps (equivalent to 37 kcps in 2D). The gains of the 3D method assessed by this analysis are significant, and are shown to be highly dependent on countrate and object dimensions

Additional details

Publishing Information

Publisher
IEEE Service Center.
Imprint Place
Piscataway, NJ (USA)
Imprint Title
IEEE nuclear science symposium conference record emdash 1990
Imprint Pagination
1636 p.
Journal Page Range
p. 1217-1221.

Conference

Title
1990 Institute of Electrical and Electronics Engineers (IEEE) nuclear science symposium.
Dates
22-27 Oct 1990.
Place
Arlington, VA (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
22090314
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BRAIN; COUNTING RATES; IMAGE PROCESSING; MEASURING METHODS; PERFORMANCE; PHANTOMS; POSITRON COMPUTED TOMOGRAPHY; THREE-DIMENSIONAL CALCULATIONS
Descriptors DEC
BODY; CENTRAL NERVOUS SYSTEM; COMPUTERIZED TOMOGRAPHY; EMISSION COMPUTED TOMOGRAPHY; MOCKUP; NERVOUS SYSTEM; ORGANS; STRUCTURAL MODELS; TOMOGRAPHY

Optional Information

Secondary number(s)
CONF-9010220--.