Published April 2016 | Version v1
Journal article

Determination of trace uranium in sea water with sodium polyphosphate as fluorescence intensifier

  • 1. Guangdong Provincial Environmental Radiation Monitoring Center, Guangzhou (China)

Description

A novel method for determination of uranium in sea water with sodium polyphosphate as fluorescence intensifier was reported. It was found experimentally that quantitative determination of trace uranium was achievable with a 10-fold dilution by deionized water, the solution pH set to 2-3 and a concentration of sodium polyphosphate of 0.58%. The relative deviation of duplicate analysis is 4.3%-6.2%, and the recovery of standard addition is 95.0%-98.2%. Its detection limit is 0.16-0.22 μg/L. (authors)

Additional details

Publishing Information

Journal Title
Radiation Protection Bulletin
Journal Volume
36
Journal Issue
2
Journal Page Range
p. 39-42
ISSN
1004-6356

Optional Information

Notes
1 fig., 7 tabs., 9 refs.