Published March 1976 | Version v1
Journal article

Optimum characteristics of a propotional counter for a detection of the cosmic X -ray radiation from 2 to 25 keV

  • 1. Nauchno-Proizvodstvennoe ob''edinenie ''Burevestnik'', Leningrad (USSR)

Description

Optimum data are given of a proportional X-radiation counter with a square cathode shape. The counter is filled up with a mixture of Xe + 10% CH4 with pressure p = 250 tor; anode diameter is 0.1 mm with an internal cathode section of 60x60 mm. These data ensure high recording efficiency, energy and time resolution in case of suppression of a high energy particle background with a discriminator with respect to acceretion fronts. Coefficient of background suppression for 60sub(Co) and 137sub(Cs) isotopes is 90% with a discriminator threshold corresponding to transmission of 70% radiation with 2-25 kev energies. The background suppression coefficient for Xe + CO2 mixture is 1.5 - 2 times lower than for the working mixture. This result does not agree with the data already published and reflects some peculiar features of squarecathode counters

Additional details

Additional titles

Original title (Russian)
Влияние радиационной загрузки на характеристики пропорциональных счетчиков

Publishing Information

Journal Title
Prib. Tekh. Ehksp.
Journal Issue
no.2
Series
Prib. Tekh. Ehksp.

Optional Information

Notes
12 refs.; for English translation see the journal Instrum. Exp. Tech.