Experimental determination of the overall perturbation factor for the NACP chamber in electron beams for dmax < d≤d80

  • 1. University of Chicago, Department of Radiation Oncology, Chicago, IL 60637 (United States).

Description

In electron beam dosimetry with an ionization chamber, a factor that corrects for the cavity perturbation of the medium, Prepl, and one to account for the disturbance due to the chamber wall material differing from the medium, Pwall, are required. The overall perturbation correction factor, pq=PreplPwall, has been introduced because of the difficulty in separately measuring these two components. An advantage of parallel-plate ionization chambers is that pq has been shown to be close to unity at dmax. However, many dosimetry applications require knowledge of the overall perturbation factor at depths greater than dmax. We determined pq for the NACP chamber at depths beyond dmax by intercomparing percentage depth dose measurements made with it with those obtained with a PTW/diamond detector for which pq was taken as unity at all the measurement depths. Data were obtained at depths corresponding to approximately the 90 and 80 per cent of the dose maxima for 20, 16, 12 and 6 MeV incident electrons. The beam energy at depth, Ed, and the percentage depth-dose gradient varied from 1.4 to 14.3 MeV and 0 to 5.8% mm-1 respectively. Our results show that within the estimated uncertainty of 1.3%, pq,NACP is unity over the range of energies and dose gradients studied. (author)

Availability note (English)

Available online at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
46
Journal Issue
2
Journal Page Range
p. N49-N55
ISSN
0031-9155

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
32007481
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
DEPTH DOSE DISTRIBUTIONS; DISTURBANCES; ELECTRON BEAMS; ELECTRON DOSIMETRY; IONIZATION CHAMBERS; RADIATION DOSES
Descriptors DEC
BEAMS; DOSES; DOSIMETRY; LEPTON BEAMS; MEASURING INSTRUMENTS; PARTICLE BEAMS; RADIATION DETECTORS; RADIATION DOSE DISTRIBUTIONS; SPATIAL DOSE DISTRIBUTIONS