Dosimetric evaluation of the Siemens Virtual Wedge
Creators
- 1. Department of Radiation Oncology, Division of Clinical Physics and Instrumentation, University Hospital Rotterdam, Daniel den Hoed Cancer Center/Dijkzigt Hospital, Rotterdam (Netherlands)
Description
Recently, Siemens has introduced its Virtual Wedge (VW). On a Mevatron accelerator, this option generates a wedge-like dose profile by moving a collimator jaw at constant speed while varying the dose rate. In this paper the formalism is given that is used to deliver a wedge profile and from that the expressions for possible combinations of wedge angle, field size and delivered MUs are derived. Also the time needed to deliver a VW field is calculated. An effective attenuation coefficient μ is used in the implementation. For three beam energies, values of μ are determined in order to get VW angles that are as close as possible to the hard wedge angles, over a wide range of field sizes and wedge angles. Linearity with number of MUs and gantry angle dependence of the generated dose profiles were checked. These factors did not have a significant influence on the VW dose profiles. Wedge factors should be close to unity in the VW implementation. We have measured a number of wedge factors and found that they start to deviate from 1 with more than 1% for large wedge angles and field sizes, up to 3.5% for a 19x19cm2, 60 deg. VW field. The Virtual Wedge turned out to be a reliable tool that can be used clinically, provided that it can be handled by the treatment planning system. It provides extra flexibility and usually results in shorter beam on times. (author)
Additional details
Publishing Information
- Journal Title
- Physics in Medicine and Biology (Online)
- Journal Volume
- 43
- Journal Issue
- 9
- Journal Page Range
- p. 2651-2663
- ISSN
- 1361-6560
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 43041974
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- ACCELERATORS; ATTENUATION; BEAM PROFILES; COLLIMATORS; DOSE RATES; MEASURING INSTRUMENTS; RADIATION DOSES; RADIOTHERAPY
- Descriptors DEC
- DOSES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY
Optional Information
- Notes
- 10 refs; This record replaces 31036353