Stereotactic radiation therapy with linear accelerator. Accuracy of alignment and portal film verification

  • 1. Nagasaki Univ. (Japan). School of Medicine

Description

Stereotactic radiation therapy with a linear accelerator requires precise alignment of therapeutic radiation distribution to the target volume. To verify the accuracy of alignment, differences between the stereotactic coordinates of the center of the therapeutic radiation distribution determined from portal films and those of the target points determined from CT/MRI or CT/angiographic localization were calculated for 58 points. The average values and one standard deviation were -0.02±0.50 mm, 0.37±0.39 mm and 0.01±0.25 mm in the x, y, and z directions, respectively. The average value of the total deviation was 0.73±0.07 mm. Potential sources of misalignment were misaligned laser pointers, wide laser beam width, improperly zeroed target positioner scale, and motion of the patient couch. The accuracy of alignment should be verified prior to irradiation. Portal film verification is indispensable for stereotactic radiation therapy with a linear accelerator. (author)

Additional details

Publishing Information

Journal Title
Nippon Igaku Hoshasen Gakkai Zasshi
Journal Volume
55
Journal Issue
8
Journal Page Range
p. 593-596.
ISSN
0048-0428
CODEN
NHGZAR

INIS

Country of Publication
Japan
Country of Input or Organization
Japan
INIS RN
27001587
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ACCURACY; DOSIMETRY; LINEAR ACCELERATORS; PLANNING; QUALITY ASSURANCE; RADIOTHERAPY; THREE-DIMENSIONAL CALCULATIONS
Descriptors DEC
ACCELERATORS; MEDICINE; THERAPY