Published October 2015 | Version v1
Miscellaneous Open

Dosimetric validation of planning system Eclipse 10 in partial breast irradiation treatments with IMRT

  • 1. Centro Medico Nacional Siglo XXI, Hospital de Oncologia, Departamentos de Fisica Medica y Radioterapia, Av. Cuauhtemoc 330, 03020 Mexico D. F. (Mexico)

Description

Partial breast irradiation is a new type of external radiation therapy to treat breast cancer in early clinical stages. Consist of administering to the channel surgical high doses of radiation in few treatment sessions. In this paper the dose calculations of the planning system Eclipse version 10 for a treatment of partial breast irradiation with X-rays beams (6 MV) intensity modulated were compared against the measurements made with OSL dosimeters and radio-chromic dye film. An anthropomorphic mannequin was used in which OSL dosimeters were collocated near the surface, an inside the radio-chromic dye film one plate; with this latest one dimensional dose distribution was measured. Previously dosimeters were calibrated irradiating them with a beam of X-rays 6 MV under the conditions specified in the IAEA-398 protocol. The OSL dosimeters were read in the Micro star Landauer equipment, the radio-chromic dye films were read with a scanner Epson 10000-Xl and analyzed with FilmCal and PTW Verisoft programs. The differences between measured and calculated dose were as follows: 3.6±1% for the OSL dosimeter and 96.3±1% of the analyzed points approved the gamma index criterion (3%, 3m m) when comparing the matrices of calculated dose and measured with the radio-chromic dye film. These results confirm the good dosimetric performance of planning system used under specific conditions used in the partial breast irradiation technique. (Author)

Files

47020861.pdf

Files (337.6 kB)

Name Size Download all
md5:a1e7156abde82cc8a55a3526ebb103ae
337.6 kB Preview Download

Additional details

Additional titles

Original title (Spanish)
Validacion dosimetrica del sistema de planeacion Eclipse 10 en tratamientos de irradiacion parcial de mama con IMRT

Publishing Information

Publisher
Sociedad Mexicana de Irradiacion y Dosimetria
Imprint Place
Mexico, D. F. (Mexico)
Imprint Pagination
10 p.
Report number
INIS-MX--2935

Conference

Title
15. International Symposium on Solid State Dosimetry
Original Conference Title
15. Conferencia Internacional sobre Dosimetria de Estado Solido
Dates
26-30 Sep 2015
Place