MiX: a position sensitive dual-phase liquid xenon detector
Creators
- 1. Randall Laboratory of Physics, University of Michigan, Ann Arbor, Michigan 48109-1040 (United States)
- 2. INPAC, Department of Physics and Astronomy and Shanghai Key Laboratory for Particle Physics and Cosmology, Shanghai Jiao Tong University, 800 Dongchuan Road, Shanghai, 200240 (China)
Description
The need for precise characterization of dual-phase xenon detectors has grown as the technology has matured into a state of high efficacy for rare event searches. The Michigan Xenon detector was constructed to study the microphysics of particle interactions in liquid xenon across a large energy range in an effort to probe aspects of radiation detection in liquid xenon. We report the design and performance of a small 3D position sensitive dual-phase liquid xenon time projection chamber with high light yield (Ly122=15.2 pe/keV at zero field), long electron lifetime (τ > 200 μs), and excellent energy resolution (σ/E = 1% for 1,333 keV gamma rays in a drift field of 200 V/cm). Liquid xenon time projection chambers with such high energy resolution may find applications not only in dark matter direct detection searches, but also in neutrinoless double beta decay experiments and other applications
Availability note (English)
Available from http://dx.doi.org/10.1088/1748-0221/10/10/P10040Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Instrumentation
- Journal Volume
- 10
- Journal Issue
- 10
- Journal Page Range
- p. P10040
- ISSN
- 1748-0221
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 47067938
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- DOUBLE BETA DECAY; ELECTRONS; ENERGY RESOLUTION; GAMMA RADIATION; LIQUEFIED GASES; NONLUMINOUS MATTER; PARTICLE INTERACTIONS; PERFORMANCE; RADIATION DETECTION; THREE-DIMENSIONAL CALCULATIONS; TIME PROJECTION CHAMBERS; XENON
- Descriptors DEC
- BETA DECAY; BETA-MINUS DECAY; DECAY; DETECTION; DRIFT CHAMBERS; ELECTROMAGNETIC RADIATION; ELEMENTARY PARTICLES; ELEMENTS; FERMIONS; FLUIDS; GASES; INTERACTIONS; IONIZING RADIATIONS; LEPTONS; LIQUIDS; MATTER; MEASURING INSTRUMENTS; MULTIWIRE PROPORTIONAL CHAMBERS; NONMETALS; NUCLEAR DECAY; PROPORTIONAL COUNTERS; RADIATION DETECTORS; RADIATIONS; RARE GASES; RESOLUTION