Analysis of translational errors in frame-based and frameless cranial radiosurgery using an anthropomorphic phantom
Creators
- 1. Faculdades Pequeno Principe (FPP), Curitiba, PR (Brazil)
- 2. Centro de Radioterapia Sao Sebastiao, Florianopolis, SC (Brazil)
- 3. Centro de Diagnostico Medico Imagem, Florianopolis, SC (Brazil)
- 4. Universidade Tecnologica Federal do Parana (UTFPR), Curitiba, PR (Brazil)
Description
Objective: To evaluate three-dimensional translational setup errors and residual errors in image-guided radiosurgery, comparing frameless and frame-based techniques, using an anthropomorphic phantom. Materials and Methods: We initially used specific phantoms for the calibration and quality control of the image-guided system. For the hidden target test, we used an Alderson Radiation Therapy (ART)-210 anthropomorphic head phantom, into which we inserted four 5- mm metal balls to simulate target treatment volumes. Computed tomography images were the taken with the head phantom properly positioned for frameless and frame-based radiosurgery. Results: For the frameless technique, the mean error magnitude was 0.22 ± 0.04 mm for setup errors and 0.14 ± 0.02 mm for residual errors, the combined uncertainty being 0.28 mm and 0.16 mm, respectively. For the frame-based technique, the mean error magnitude was 0.73 ± 0.14 mm for setup errors and 0.31 ± 0.04 mm for residual errors, the combined uncertainty being 1.15 mm and 0.63 mm, respectively. Conclusion: The mean values, standard deviations, and combined uncertainties showed no evidence of a significant differences between the two techniques when the head phantom ART-210 was used. (author)
Availability note (English)
Available from http://www.rb.org.br/audiencia_pdf.asp?aid2=2713&nomeArquivo=en_v49n2a10.pdfAdditional details
Publishing Information
- Journal Title
- Radiologia Brasileira
- Journal Volume
- 49
- Journal Issue
- 2
- Journal Page Range
- p. 98-103
- ISSN
- 0100-3984
- CODEN
- RDBRAS
INIS
- Country of Publication
- Brazil
- Country of Input or Organization
- Brazil
- INIS RN
- 47065358
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- CALIBRATION; COMPUTERIZED TOMOGRAPHY; ERRORS; HEAD; IMAGES; PHANTOMS; QUALITY CONTROL; RADIOTHERAPY; SKULL; SURGERY; THREE-DIMENSIONAL CALCULATIONS
- Descriptors DEC
- BODY; CONTROL; DIAGNOSTIC TECHNIQUES; MEDICINE; MOCKUP; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; SKELETON; STRUCTURAL MODELS; THERAPY; TOMOGRAPHY