Published October 2003
| Version v1
Journal article
Design and development of an interchangeable nanomicroelectrospray source for a quadrupole mass spectrometer
- 1. Department of Chemistry, Natural Sciences Complex, University at Buffalo - State University of New York, Buffalo, New York 14260-3000 (United States)
Description
An interchangeable microelectrospray and nanoelectrospray ionization source has been designed and constructed for use on a triple-quadrupole mass spectrometer. This new source is specially designed to conduct nanoelectrospray experiments utilizing pulled borosilicate glass emitters, as well as various ionization modes utilizing silica capillaries. The source design facilitates the easy exchange between a microelectrospray source and nanoelectrospray source. The microionspray assembly can be quickly replaced by the nanoelectrospray adaptation in order to conduct nanoelectrospray experiments. Furthermore, this source design allows for the coupling of low flow separation techniques, like microliquid chromatography or capillary electrophoresis
Additional details
Identifiers
- DOI
- 10.1063/1.1611994;
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 74
- Journal Issue
- 10
- Journal Page Range
- p. 4474-4477
- ISSN
- 0034-6748
- CODEN
- RSINAK
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 36005623
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BOROSILICATE GLASS; CAPILLARIES; CHROMATOGRAPHY; COUPLING; COUPLINGS; DESIGN; ELECTROPHORESIS; ION SOURCES; IONIZATION; MASS SPECTROMETERS; SILICA
- Descriptors DEC
- BLOOD VESSELS; BODY; CARDIOVASCULAR SYSTEM; GLASS; MEASURING INSTRUMENTS; MINERALS; ORGANS; OXIDE MINERALS; SEPARATION PROCESSES; SPECTROMETERS
Optional Information
- Notes
- (c) 2003 American Institute of Physics.