Published October 2003 | Version v1
Journal article

Design and development of an interchangeable nanomicroelectrospray source for a quadrupole mass spectrometer

  • 1. Department of Chemistry, Natural Sciences Complex, University at Buffalo - State University of New York, Buffalo, New York 14260-3000 (United States)

Description

An interchangeable microelectrospray and nanoelectrospray ionization source has been designed and constructed for use on a triple-quadrupole mass spectrometer. This new source is specially designed to conduct nanoelectrospray experiments utilizing pulled borosilicate glass emitters, as well as various ionization modes utilizing silica capillaries. The source design facilitates the easy exchange between a microelectrospray source and nanoelectrospray source. The microionspray assembly can be quickly replaced by the nanoelectrospray adaptation in order to conduct nanoelectrospray experiments. Furthermore, this source design allows for the coupling of low flow separation techniques, like microliquid chromatography or capillary electrophoresis

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
74
Journal Issue
10
Journal Page Range
p. 4474-4477
ISSN
0034-6748
CODEN
RSINAK

INIS

Optional Information

Notes
(c) 2003 American Institute of Physics.